Physics of ultrasound
Wavelength, frequency, resolution — what drives image quality.
6 min read
Learning objectives
- You know the basic quantities of a sound wave (frequency, wavelength, amplitude).
- You understand the trade-off between penetration and resolution.
- You can name typical transducer frequencies for adult TTE.
The physics of ultrasound is the foundation of every echocardiographic exam: to optimise image quality on purpose, you have to understand how frequency, wavelength and penetration depth are related.
The basic quantities
A sound wave is a periodic pressure fluctuation, described by:
- Frequency f — oscillations per second, in Hz.
- Wavelength λ — distance between two maxima, in mm.
- Amplitude A — peak pressure deviation; determines intensity.
- Speed of sound c — roughly constant in soft tissue at ≈ 1540 m/s.
The relation is c = f · λ, hence λ = c / f.
Frequency vs. penetration
| Frequency | Penetration | Axial resolution |
|---|---|---|
| 2 MHz | ~ 30 cm | coarse |
| 3.5 MHz | ~ 20 cm | medium |
| 5 MHz | ~ 12 cm | good |
| 7.5 MHz | ~ 6 cm | high |
Trade-off: higher frequency means better resolution, but stronger attenuation. Adult TTE typically uses 2–3.5 MHz, TEE 5–7 MHz, paediatric probes 5–8 MHz.
Resolution
- Axial resolution ≈ ½ pulse length → improves with higher frequency.
- Lateral resolution depends on beam width → varies with focus.
- Temporal resolution (frame rate) trades off against sector size and line density.
Mini-quiz
1 / 2How are frequency and wavelength related in soft tissue?
Take-aways
- Higher frequencies give better axial resolution but less penetration.
- Speed of sound in soft tissue: c ≈ 1540 m/s.
- Adult TTE probes typically operate at 2–5 MHz; paediatric/TEE probes are higher.